Sarns TCM II Heater Cooler

The Sarns TCM II is a computer controlled heating and cooling system for use during cardiopulmonary bypass. These units are pre-owned and in excellent condition. Please contact us for pricing information.

SKU: ES0010 Categories: ,

Description

The Sarns TCM II is a computer controlled heating and cooling system for use during cardiopulmonary bypass.  The basic features include the following:

  • Independent ice maker
  • Precise temperature control
  • Oxygenator and cardioplegia water supply
  • Optional remote control and display module
  • More Informationn from Terumo

Specifications

 

-
Electrical
-
 44154416
Voltage1008/122V (115V nominal)210/250V (220/240V nominal)
Frequency60 Hz50Hz
Amperage16 amps12 amps
Current leakage100 microamps maximum for 115V500 microamps maximum for 220V
Ground resistanceMax. 0.1 OhmMax. 0.1 Ohm
-
Dimensions
-
 Main Unit
Height34.4″  (87.4 cm)
Width20.0″  (50.8 cm)
Depth31.5″ (80.0 cm)
Weight282 lbs (127.9 kg) with tanks empty
356 lbs (161.5 kg) with tanks full
 Large TankSmall Tank
Length18.5″ (47.0 cm)9.5″ (24.1 cm)
Width8.7 (22.1 cm)5.5″ (14.0 cm)
Depth21.5″ (54.6 cm)9.3″ (23.6 cm)
Capacity: Water
Capacity: Crushed ice
9.0 gal (34 L)
40.0 lbs (18.1 kg)
1.2 gal (4.5 L)
7.5 lbs (3.4 kg)
 Temperature Display UnitTemperature Probe Modules
Length8.75″ (22.2 cm)9.0″ (22.9 cm)
Width5.5″ (14.0 cm)5.75″ (14.6 cm)
Depth2.5″ (6.4 cm)2.5″ (6.4 cm)
-
Water Outlet Temperature
-
Arterial heat exchange and blanket0 – 42° C
CardioplegiaIce water only
Heaters1 each – 700 watt, 850 watt, 115V units
2 each – 1000 watt, 220V units
-
Maximum Flow Performance (1/8 hp pump)
-
Arterial heat exchanger* outlet port5.5 gal/min (20.8 L/min) at zero head
13.5 psi (698 mmHg) maximum pressure
Blanket supply outlet port*0.8 gal/min (3.0 L/min) at zero head
11.8 psi (610 mmHg) maximum pressure
Cardioplegia water supply port2.1 gal/min (7.9 L/min)
7.5 psi (388 mmHg) maximum pressure
*Note: performance will be different with both a heat exchanger and a blanket being supplied at the same time. 
-
Ice Supply
-
If the large tank is filled to the maximum level with 25° C water, the ice maker can provide a maximum of 50.0 lbs (22.7 kg) in 8 hours.
If the large tank is filled to the maximum level with 3° C water, the ice maker can provide a maximum of 33.5 lbs (15.1 kg) in 3 hours.
-
Temperature Probe Module Operating Range
-
0 – 50°C
-
Remote Temperature Sensing and Display Accuracy
-
 ±0.3°C
-
Venous Differential Mode
-
6°C gradient = rewarm water temperature is held to a maximum of 6 76°C more than the incoming venous blood temperature.
10°C gradient = rewarm water temperature is held to a maximum of 10°C more than the incoming venous blood temperature.

Additional information

Weight282 lbs